SET UP MODE

The set up mode is used
This set up mode may only be entered by healthcare professionals and authorised personnel trained to use this application of
the pump.
The patient or his/her relatives without permission of the physician, dietician, nurse or other licensed practitioner, may not
change the settings of the set up mode.
+
+
&
2 sec
2 sec
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode.
First the alarm level can be set: "BEEP HIGH" (+) or "BEEP LOW" (-)
Use the "+" or "-" key to change the setting from "HIGH" to "LOW" or from "LOW" to "HIGH".
+
LOCK
fill set
UNLK
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to switch to the next setting:
"UNLK" the keypad (-) or "LOCK" the keypad (+). In the "LOCK" mode the feeding program of the pump can no longer be changed.
Use the "+" or "-" key to change the setting from "LOCK" to "UNLK" or vice versa.
+
BEEP WHEN DONE
fill set
MUTE WHEN DONE
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to switch to the next setting:
"BEEP WHEN DONE" (+): the pump will give an alarm when the dose is administered or "MUTE WHEN DONE" (-): the pump will not give an alarm after
administering the dose. Use the "+" or "-" key to change the setting from "BEEP WHEN DONE" to "MUTE WHEN DONE" or vice versa.
+
LITE ON
fill set
LITE OFF
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to switch to the next setting:
ZMNL900907_AW.indd 28
ZMNL900907_AW.indd 28
UK
• to set the alarm level,
• to lock the keypad,
• to switch the audible alarm off when a dose is administered.
• to switch the light of the LCD permanently on when connected to the mains.
• to enable connection with a Nurse call or Patient Data Monitoring System
(PDMS).
TO SET THE ALARM LEVEL
BEEP HIGH
BEEP LOW
TO LOCK THE KEYPAD
TO MUTE WHEN DOSE DONE
TO SWITCH LIGHT ON
-26-
"