Operation Mode - Minipa MT-360 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

meter is turned on. To change the temperature scale
by pressing °C / °F key.
APO Key
It will auto power off for about 10s.
Press APO key to disable Auto Power Off function
that HOLD indication disappears and press again to
enable APO function.

5.2 Operation Mode

1. Take the protective cap off and them pull the trig-
ger to turn on the meter.
2. Point the lens at the object whose temperature is
to be measured.
3. Measurement is performed as long as trigger is
kept.
4. Referring to the spot size figure, aim the laser
beam at the object whose temperature is to be
measured.
5. Pull the cap on to extend life of the sensor and to
avoid danger caused by wrong way to use laser.
REMARK: Although the field of measurement (or
field of view) and the spot almost coincide, actu-
ally the field of measurement corresponds to the
diameter for 90% optical response. The object
whose temperature is to be measured needs to
be larger than the measurement diameter (spot of
size) by an adequate margin at least 1.5 to 2 times
larger.