5.4.2

Machine Cleaning and Disinfection

Automated cleaning and disinfection is not possible.
5.5

Inspection, Functional Check, Maintenance

5.5.1

Inspection and Functional Check

Further use is confirmed by the successful inspection of the product.
• The products must be macroscopically clean, i.e. free from visible contamination, after each
cleaning process.
• Inspect the products for fractures, cracks, deformation, damage and proper function.
• Worn, deformed, porous or otherwise damaged products must be replaced.
5.5.2

Maintenance

n/a
5.6

Packaging

n/a
5.7

Sterilization

Cannot be sterilized.
5.8

Storage and Transport

• Store and transport cleanly.
• Avoid severe temperature fluctuations.
• Protect the products against mechanical damage.
• If the headlight is to be stored for more than two weeks, remove the batteries and place them in the
original packaging.
• Store batteries only at ambient temperatures from 0°C to +25°C (+32°F to +77°F).
• Use the charging unit only in dry environments in a closed area. Protect from rain, snow, inflammable
materials, excessive heat and/or moisture.
Where applicable, the specific parameters for ambient conditions during the application (temperature
range and air humidity) as stated on the product label must be observed.
46