Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain
a device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to
maintain a device may compromise the clinical condition or safety of users and/or their caregivers. Do not
perform service and maintenance while the user is in the product.
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any
damaged or worn-out parts
Ensure that all fasteners are in place and fitted
correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Weekly
Check that all castors are moving freely and
that all wheel locks/brakes are easily activated
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild
detergent/soap and dry before use
Check that the angle mechanism functions
correctly and that the safety lock is working
correctly
Disinfect the product. please check the
disinfection section in this User Guide for more
information
Monthly
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substandard
or faulty
support.R82.org
Warewashing machine
This product can be washed at:
60°C in 10 minutes
85°C in 3 minutes
.. using a mild detergent in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories separately.
Before washing
Remove the following items and clean them
separately:
Accessories
Cushioning
Spoke protectors
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to wipe
off any residue and dirt from the product, using a
cloth with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in
accordance with the manufacturer's instructions
and demonstrably recorded.
Service Check List:
Check and adjust the following functions as per their
intended use. Repair or replace any damaged or
worn-out parts. Please check the:
Wheels are moving freely and that all wheel
locks/brakes are easily activated. Adjust if
required
Spoke guards is without cracks and secured
to the wheel with the clips mounted on each
guard
Castors are moving freely and all wheel locks/
brakes are easily activated
The column angle is adjustable and it is posible
to lock it in position
Nuts and bolts on the product are tightened
Oil the swing-away parts
Check all fittings and buckles for signs of worn
out parts
Product for cracks or signs for worn-out parts
All labels on the product are intact