Capanivo
Conexión eléctrica
FM / CSA Approval - Connection drawing
FM / CSA Class I, II, III Div.1, Group A-G
Enclosure Ø65mm
Enclosure Ø35mm
For use at altitude ≤ 2000m:
Ambient temperature
range
-40 .. +50°C (-40 .. +122°F) (1)
-40 .. +65°C (-40 .. +149°F) (1)
-40 .. +85°C (-40 .. +185°F) (1)
-40 .. +85°C (-40 .. +185°F) (1)
(1) With option FFKM O-ring seal: Lower ambient and process temperature limited to -20°C (-4°F)
For use at altitude > 2000m ... ≤ 3000m:
Ambient temperature
range
-40 .. +45°C (-40 .. +113°F) (1)
-40 .. +58°C (-40 .. +136°F) (1)
-40 .. +76°C (-40 .. +168°F) (1)
-40 .. +76°C (-40 .. +168°F) (1)
(1) With option FFKM O-ring seal: Lower ambient and process temperature limited to -20°C (-4°F)
CN 7000
Interruptor de nivel
®
Serie CN 7000 - Relé DC y 8/16mA
Información técnica / Manual de instrucciones
HAZARDOUS LOCATION
Power supply /
8/16mA loop
Solid state relay
(signal output),
optional
Connect with
equipotential
bonding of the plant.
8/16mA loop
Pin 2,4 not used
Process temperature
range
-40 .. +50°C (-40 .. +122°F) (1)
-40 .. +65°C (-40 .. +149°F) (1)
-40 .. +100°C (-40 .. +212°F) (1)
-40 .. +125°C (-40 .. +257°F) (1)
Process temperature
range
-40 .. +45°C (-40 .. +113°F) (1)
-40 .. +58°C (-40 .. +136°F) (1)
-40 .. +90°C (-40 .. +194°F) (1)
-40 .. +112°C (-40 .. +233°F) (1)
NON HAZARDOUS LOCATION
Associated
apparatus
IS
non
IS
IS
non
IS
The power supply circuit is infallibly galvanically
isolated from the solid state relay circuit.
Use supply cables rated at least 90°C.
IS
non
IS
Use supply cables rated at least 90°C.
Max. Surface
temperature
80°C
95°C
130°C
155°C
Max. Surface
temperature
80°C
95°C
130°C
155°C
gi060821
Barrier or Signal
Conditioning
Instrument
Switch amplifier
for contact input
Signal
Conditioning
Instrument
Temperature
class
T6
T5
T4
T3
Temperature
class
T6
T5
T4
T3
Pág.
1
2
3
4
5
6
7
8
9
10
11
12
27