Saving Data; Outputting The Data Memory; Transmitting Data (Only Ph 340I) - wtw pH 330i Instrucciones De Operación

Ocultar thumbs Ver también para pH 330i:
Idiomas disponibles

Idiomas disponibles

pH 330i/340i

Saving data

manually
Saving data
automatically
Int 1
Transmitting data
manually
Transmitting data
automatically
Int 2
All manuals and user guides at all-guides.com
Saving data
Press the <STO> key in the measuring mode (display No. with the number of
the next free memory location). Then, press <RUN/ENTER> and enter the ID
number with < > < >. Terminate the save with <RUN/ENTER>. The instru-
ment changes to the measuring mode.
The portable measuring instrument pH 330i/340i has an internal data mem-
ory. It can store up to 800 datasets. The save interval (Int 1) determines the
chronological interval between automatic save processes.
To set up the save interval, press <STO> while pressing the <RUN/ENTER>
key (display Int 1) and set the interval with < > < > key. Then, press
<RUN/ENTER> and enter the ID number with < > < >. Change to the last
active measuring mode with <RUN/ENTER>. The automatic save is switched
on (display Auto Store).

Outputting the data memory

You can output the data memory with the <RCL> key. By repeatedly press-
ing the <RCL> key, you reach the following functions:
StO dISP
Output stored data on the display
StO SEr
Output stored data via the serial interface (only pH 340i)
CAL dISP
Output calibration data on the display
CAL SEr
Output calibration data via the serial interface (only pH 340i)
Start the output with <RUN/ENTER>.

Transmitting data (only pH 340i)

Press the <RUN/ENTER> key in the measuring mode.
This manually triggers a data transmission of the current measured value to
the serial interface at any time - independently of the selected intervals.
The interval to the data transmission (Int 2) determines the chronological in-
terval between automatic data transmissions. After the selected interval ex-
pires, the current data record is transmitted to the interface. To set up the
transmission interval, press <RCL> while holding down the <RUN/ENTER>
key (display Int 2). Then, set the interval with < > < >.
Saving data
11
loading

Este manual también es adecuado para:

Ph 340i