Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain
a device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to
maintain a device may compromise the clinical condition or safety of users and/or their caregivers. Do not
perform service and maintenance while the user is in the product.
1st Day
EN
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any
damaged or worn-out parts
Ensure that all fasteners are in place and fitted
correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Check that all cables connecting the power
system are in good shape
Weekly
Check that all castors are moving freely and
that all wheel locks/brakes are easily activated
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild
detergent/soap and dry before use
Disinfect the product. please check the
disinfection section in this User Guide for more
information
Monthly
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substandard
or faulty
Warewashing machine
This product can be washed at:
60°C in 10 minutes
85°C in 3 minutes
.. using a mild detergent in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories separately.
Before washing
Remove the following items and clean them
separately:
Accessories
Power unit including; Actuator, battery and
handset
After washing:
Replace the power column going backwards
through the procedure give on page 14
Check that the handset functionalities works
correct
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to wipe
off any residue and dirt from the product, using a
cloth with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in
accordance with the manufacturer's instructions
and demonstrably recorded.
Service Check List:
Check and adjust the following functions as per their
intended use. Repair or replace any damaged or
worn-out parts. Please check the:
For products with gasspring and/or telescopic
tube, check that the locking mechanisms work.
Also check that it is possible to tilt the basis
plate horizontally and vertically.
For products with power, recharge the battery
and check that the handset works by tilting the
basis plate to vertical and horizontal position
Check that all castors are moving freely and
that all wheel locks/brakes are easily activated
The foot plate(s) is adjustable, both distance
to basisplate and the angle mechanism must
be checked
All cushions are in good shape without cracks
and or holes in the cover
Nuts and bolts on the product are tightened
The straps must be without cracks and not
flossed, pay special attention to the area where
the straps connect to the brackets.
The plastic pieces connecting the frame to the
basis plate is undamaged and without cracks.
Pay special attention to the areas where
fasteners are inserted
The rails in the basis plate must be in good
shape, it should be possible to slide a strap up
and down the rail.
All labels on the product are intact