CAUTION
Do not allow any foreign substances or liquid substances,
such as dirt, blood, or water, enter into the meter. The meter
may be damaged or may malfunction. Follow the warning
information provided below to prevent possible damage to
the meter.
• Do not apply the blood or control solution samples directly
to the test strip port.
• Do not apply the blood or control solution samples to the
test strip while holding the meter in a way that the tip of
the test strip faces upwards. The blood or control solution
samples may run down the surface of the test strip and flow
into the test strip port.
• Do not store your meter in unsanitary or contaminated sites.
• Make sure to follow the Pre-cleaning and Disinfection
Procedures found in the Caring for the ReliOn™ Premier
Classic System section of this user manual.
NOTE
• You can get more safety information at FDA Public Health
Notification (http://wayback.archive-it.org/7993/
20170111013014/http://www.fda.gov/MedicalDevices/
Safety/AlertsandNotices/ucm224025.htm) or at CDC Clinical
Reminder (www.cdc.gov/injectionsafety/Fingerstick-
DevicesBGM.html).