Switching On/Off The Temperature Compensation (Tc) - wtw Cond 3150i Instrucciones De Operación

Idiomas disponibles

Idiomas disponibles

Cond 3150i
ips
Application tips
t
c
2
3 8
S/
cm
n
L
F
Tref25
ARng
nLF
All manuals and user guides at all-guides.com
Switching on/off the temperature compensation
(TC)
The basis for calculating the temperature compensation is the selected refer-
ence temperature, Tref 25 or Tref 20. For details of how to select the refer-
ence temperature, see C
You can select from the following settings:
Nonlinear temperature compensation (nLF) according to EN 27 888
No temperature compensation (- - - -)
Note
Select the following temperature compensations given in the table according
to the respective test sample:
Sample
Natural water (ground
water, surface water and
drinking water)
Ultrapure water
Other aqueous solutions
Salinity
(seawater)
Procedure for switching temperature compensation on/off:
Press the <CAL> key repeatedly until
Then press the <RUN/ENTER> key.
Switch the temperature compensation on/off with the <CAL> key.
Selection: Nonlinear temperature compensation (nLF), no temperature
compensation (- - - -).
Switch to the measuring mode with <M>.

Switching on/off the temperature compensation (TC)

.
ONFIGURATION
Temperature compensation
nLF
according to DIN 38404
EN 27 888
nLF
according to DIN 38404
EN 27 888
Measure without temperature
compensation
Automatically nLF according to
IOT
tc appears on the display.
Display
indicator
- - - -
Sal,
7
loading