User manual
12. Service Menu
The instrument has a password protected service menu. To enter the
service menu press the touch-screen three times during the self test after
turning the equipment on. Upon request input the PIN "1234".
A selective list with different menu items appears.
Display 24 : Service menu
12.1 How to reset the system (Load Default)
Select 'Load default' from the service menu. The instrument will be reset to
delivery status. All settings modified by the user as well as the memory will
be cleared!
12.2 How to control the LOT-control (LOT activate)
Select 'LOT activate' from the service menu to activate the LOT-control.
This will minimize the risk of using expired test strips. You will be asked for
the LOT of strips currently used. The system will give a warning when the
test strips are expired. After 100 measurements the system will ask for the
LOT-number of the next tube.
Manual EN V2.00 / 06-2009
41