Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
portable and mobile RF communications equipment and the [PO3M]
The PO3M is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the PO3M can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the PO3M as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output
power of
transmitter
W
0,01
0,1
1
10
100
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
d=1.2 p
0.12
0.38
1.2
3.8
12
m
d=1.2 p
0.12
0.38
1.2
3.8
12
19
d=2.3 p
0.23
0.73
2.3
7.3
23