WAVEFORMS (CONTINUED)
TENS- Symmetrical Biphasic
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
The Symmetrical Biphasic waveform has a short pulse duration
and is capable of strong stimulation of nerve fibers in the skin
and in muscle. This waveform is often used in portable muscle
stimulation units, and some TENS devices.
Output Mode
Available on Channel
Treatment Time (Stim)
Treatment Time (Combo)
Mode Selection (Stim)
Mode Selection (Combo)
Output Intensity
Amplitude Modulation
Burst Frequency
Cycle Time
Frequency
Frequency Sweep
Phase Duration
Ramp
Sweep Time
Sweep Low Frequency
Sweep High Frequency
IRMS
DC component
Electrodes
1,2
1-60 min
1-30 minutes
CC or CV
CV
0-140 mA (CC) 0-140 V (CV)
0% (off) to 100% on 10% steps
0-10 bps
Continuous or User Defined
1-200 pps
On/Off
30-400 µsec
0-5 sec
14sec
1-199 pps
2-200 pps
0-50mA
No
TENS - HAN
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
The HAN Waveform provides optimal parameters with a precisely
controlled sequence of Dense-and-Disperse (DD) modes of
stimulation where a burst of 8 pulses at 80Hz is alternating with
continuous stimulation (no burst), each lasting for 3 seconds. This
produces a synergistic analgesic effect.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Burst Frequency
Frequency
Phase Duration
IRMS
DC component
E N
Electrodes
1, 2
1-60 min
CC
0-100 mA (CC)
2 bps
80 pps
180 μsec
0-19mA
No