Retrieve Switch Test Results; Running The Tecal 700Xs From A Pc; Connecting The Unit To The Computer Using A Usb Cable - Techne TECAL 700XS Manual Del Operador

Tabla de contenido
P 1500
set
E 1801
set
Contacts Closed
P 1800
Contacts Open
P 2000
The Tecal 700XS can be run from a PC using our TechneWorks software.
This software is available free of charge to download from the Techne web-sites at www.techne.com or
www.techneusa.com

Connecting the unit to the computer using a uSB cable

When connecting The Tecal 700XS to a PC for the first time using USB connection the next available
comms port will be allocated.
The hardware wizard will direct you through the installation for a USB to UART connection.
Techneworks includes a help file giving full instructions for use with the Tecal 700XS.
Test
Test

runnIng ThE TECAL 700XS from A pC

26
Exit Switch Test program
When in the Switch Test mode the test run
may be stopped at any point by pressing
the Set button. The unit will then return to
the normal set point temperature.

rETrIEVE SWITCh TEST rESuLTS

Switch Test results
Once the test results have been taken
they can be retrieved by pressing the Test
button which will show the last result. This
data is stored until the next Switch Test run
has been completed.
Switch Contact Status
When in the Switch Test mode the Set LED
shows the switch contact status. If the
switch contacts are normally closed the
LED will be on. If the contacts are normally
open the LED will be off.