End point
E2000
set
Adjust (End point)
E2000
set
P 1500
set
SWITCh TEST -End point
To check the End point for the final set
temperature of the switch test. Press the
Set button four times. (If checking after
adjusting the ramp rate press the Set
button once). End point values must be
Test
higher than the Plateau set point values
and can be set to 700°C.
Switch Test - Adjust End point
To Adjust the End point temperature press
the Set button four times and continue to
hold until the LED shown illuminates.
Then using just the up or doWn ArroW
adjust the End point value. The End point
value must be higher than the Plateau set
Test
point.
note
Values set for all of the above will be
retained after the unit is powered off.
run SWITCh TEST
Start Switch Test program
Once the P, r , E values have been set the
Switch test can be started by pressing and
holding the Test button for a few seconds.
The display will show a flashing P and the
current block temperature which will flash
Test
until the Plateau temperature is reached.
When the Plateau temperature has
stabilised the P will display continuously.
Then use the up ArroW button to
proceed to the next temperature with the
controlled heating ramp rate.
24