Seko TEKNA EVO TPR Serie Manual De Instalación página 10

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23
Potential Redox Calibration Menu (O.R.P.)
Pressing the CAL key for 3 seconds takes you into the calibration menu. If calibration was excluded during
programming, the following appears on the display:
Calibration
Off
If calibration is active:
Cal
Calibration
Automatic
It is possible to select automatic or manual mode.
Automatic calibration:
-
The buffer solution value appears on the display. Insert the probe in the bottle and press the
second countdown necessary to complete calibration will appear on the display. If the alignment quality is below
50%, an error message appears on the display and you should press
automatically after 4 seconds). If the quality is above 50%, the value is shown on the display and you should press
the
key to complete the procedure.
Manual calibration:
-
The buffer solution value appears on the display. Insert the probe in the bottle and press the
of 465 mV should now flash on the display. Insert the probe in your solution and use the
the value of the solution in your possession, then confirm by pressing the
procedure as before
EM00136063
Calibration
Automatic
-
Calibration
Manual
rev. 2.2 EN
Automatic Cal.
Buffer Sol. 465mV
Quality
||||||
-
Automatic Cal.
Buffer Sol. 465mV
Wait....
60s
Buffer Sol. 465mV
100%
Automatic Cal.
ok 465mV
Error
Wait....
+
Buffer Sol. 600mV
Quality
100%
Automatic Cal.
||||||
ok
600mV
Error
to exit calibration (the pump exits
key. The value
keys to display
key and begin the calibration
465mV
60s
600mV
key. A 60
9
Tabla de contenido
loading

Este manual también es adecuado para:

Tekna evo tpr 800Tekna evo tpr 600Tekna evo tpr 603

Tabla de contenido