7305-001-700 Rev-A
WARNING: ALWAYS sterilize the contaminated
transfer shield before using it again.
4. Remove the contaminated transfer shield.
Sterile Assistant
WARNINGS:
▪ When closing the door, hold the door by the latch.
DO NOT touch the edges of the door or the internal
surface of the door that faces the power module.
▪ ALWAYS verify the door is securely closed to prevent
exposing the power module to the sterile field.
1. Holding the door by the latch, use the door to push the
power module handle down until it locks into position
against the power module.
www.stryker.com
2. Close the door and turn the latch to the lock position.
Verify the mark on the latch aligns with the lock
symbol and the door is securely closed.
NOTE: Although one power module is usually sufficient
for one operation, Stryker recommends preparing a
second handpiece with power module to ensure quick
intraoperative exchange of handpieces under sterile
conditions.
To Install an Attachment
1. Align the J-notch with the handpiece and insert the
attachment until it snaps securely into place.
2. Gently pull on the attachment to verify it is secure.