Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of device should assure
that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
Electrical
fast
transient/
burst
IEC 61000-
4-4
Surge
IEC 61000-
4-5
IEC 60601test
level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15
kV air
± 2 kV for
power supply
lines ± 1 kV
for input/
output lines
± 0.5kV, ± 1
kV line(s) to
lines ± 0.5kV,
± 1 kV, ± 2
kV line(s) to
earth
Compliance
level
Floors should be
wood, concrete or
±8 kV
ceramic tie. If floors
contact
are covered with
±2 kV, ±4
synthetic material,
kV, ±8 kV,
the relative humidity
±15 kV air
should be at least
30%.
Mains power quality
should be that of a
Not
typical commercial
applicable
or hospital
environment.
Mains power quality
should be that of a
Not
typical commercial
applicable
or hospital
environment.
39
ENGLISH
Electromagnetic
environment -
guidance