Technical Data
AutoRep
™
S with Encode syringes, 20°C 'Ex'
Nominal
volume
Encode
syringe
0.1 mL
0.5 mL
1.0 mL
1.25 mL
2.5 mL
5.0 mL
100 – 500
10.0 mL
200 – 1000
12.5 mL
250 – 1250
25.0 mL
500 – 2500
50.0 mL
1000 – 5000
A* = Accuracy, CV* = Coefficient of variation
Error limits refer to the partial volume set relative to the Encode syringe, obtained when instrument,
syringe and distilled water are equilibrated at ambient temperature and with smooth operation. The
testing is according to DIN EN ISO 8655-5.
Important!
The AutoRep
™
S can also be operated with compatible tips from other
manufacturers. The user must check their compatibility and perform a
function and volume test before use!
Volume
range (µl)
1 ≈ 2%
2 – 10
8.0
10 – 50
4.0
20 – 100
4.0
25 – 125
4.0
50 – 250
3.5
2.5
1.5
1.5
1.5
1.5
A*
± %
<
–
Stroke setting ≈ %
of nominal volume
3 ≈ 6%
5 ≈ 10%
2.7
1.6
1.33
0.8
1.33
0.8
1.33
0.8
1.17
0.7
0.83
0.5
0.5
0.4
0.5
0.3
0.5
0.3
0.5
0.3
Accuracy Table
CV*
%
<
–
Stroke setting ≈ %
of nominal volume
1 ≈ 2%
3 ≈ 6%
5 ≈ 10%
5.0
3.0
1.4
0.73
1.0
0.38
0.8
0.38
0.8
0.3
0.6
0.27
0.5
0.23
0.3
0.23
0.4
0.23
0.4
0.23
2.0
0.6
0.4
0.3
0.2
0.2
0.2
0.2
0.2
0.15
15