Troubleshooting Module Installation
1.
Power off the oscilloscope.
2.
Remove the module and examine the contacts for damage or
contamination. Clean contacts if necessary using a 75% isopropyl
alcohol solution.
3.
Reinstall the module into the oscilloscope.
4.
Power on the oscilloscope, and perform the Checking Module
Installation procedure. (See page 6.)
a.
b. If the Application Modules list does not show the module as
5.
Power off the oscilloscope and move the module to another slot.
6.
Power on the oscilloscope, and perform the Checking Module
Installation procedure. (See page 6.)
a.
b. If the oscilloscope does not list the module after you try the
DPO2000 and MSO2000 Series Oscilloscopes
If the Application Modules list shows the module as installed,
cleaning the contacts fixed the problem.
installed, go to step 5.
If the Application Modules list shows the module as installed,
you have a problem with one of the module slots. Contact your
Tektronix service center.
module in every slot, contact your Tektronix service center.
Checking Module Installation
7