R82 Flamingo Frame Manual Del Usuario página 25

Tabla de contenido
Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device's warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers. Do not perform service and maintenance while the user is in the product.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
• Store the enclosed tools on/at the product
After every use
• Wipe the product dry
Daily
• Use a dry cloth to clean the product
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
• Check all fittings and buckles for signs of
worn out parts
• Be aware of any MAX mark indications
before adjustment
Weekly
• Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
• Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
Monthly
• Nuts and bolts on the product are tightened
• Oil the swing-away parts
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
support.R82.org
Disinfection
The product without accessories can be
disinfected with a 70% disinfectant IPA solution.
It is recommended to wipe off any residue
and dirt from the product, using a cloth with
warm water and a mild detergent/soap without
chlorine and let it dry before disinfection.
Warewashing machine
This product can be washed at 60° using a mild
detergent for 10 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
This product can be washed at 85° using a mild
detergent for 3 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
Remove the following items and clean them
separately: Cushioning and Seat
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded.
Service Check List:
Check and adjust the following functions as
per their intended use. Repair or replace any
damaged or worn-out parts. Please check the:
• Product can be folded for storage or
transportation
• Frame is adjustable in height
• Seat plane is adjustable and can be locked
in position
• Adjustment options for the foot rest, it is
possible to lock the angle and fasten the
telescopic tubes in position
• Nuts and bolts on the product are tightened
• Oil the swing-away parts
25
EN
Tabla de contenido
loading

Tabla de contenido