OHAUS a-AB23PH Guia De Inicio Rapido
Ocultar thumbs Ver también para a-AB23PH:
Idiomas disponibles

Idiomas disponibles

Enlaces rápidos

Switch On/Off
On
Main screen
Off
Buffer Group Setup
Select the buffer group
Note: Predefined buffer group 1) 1.68 4.01 7.00 10.01 12.45
Calibration
Insert a probe into the 1st
Calibrating (1
buffer solution.
Measurement
Insert the probe into a sample.
Switch pH/mV Mode
pH mode
Confirm
2) 1.68 4.01 6.86 9.18 12.45 (25°C)
st
point)
Completed
Measuring
Completed
Switch Endpoint Mode
mV mode
Auto end point mode
Temperature Setup
(
/
°C
°F)
Setup temperature unit
Completed (2
nd
point) completed
Store/Recall Data
Store measurement data:
After the measurement is
completed, press the Store
button.
Continuous reading mode
Confirm
Confirm
Setup MTC temperature
nd
Note: after the 2
calibration point is
completed, press the Store button to end
the calibration or follow the previous steps to
rd
perform a 3
point calibration.
Recall measurement data:
Recall calibration data:
press and hold the Store
After the calibration is
button to recall.
completed, press and hold
the Cal button to recall.
loading

Resumen de contenidos para OHAUS a-AB23PH

  • Página 3 Encender / Apagar Cambiar Modo pH/mV Cambiar Modo de Punto Final Encendido Modo de punto final automático Modo de lectura continua Pantalla principal Modo pH Modo mV Apagado Configuración del Grupo de Búfer Configuración de Temperatura Seleccione el grupo de búfer Confirmar Confirmar Confirmar...