5. Dispense spray for 1-2 seconds, then run handpiece to purge cleaner/lubricant.
6. Repeat steps 4 and 5 until cleaner/lubricant expelled from head is clear.
7. Wipe excess cleaner/lubricant from handpiece with dry, lint-free cloth.
Fig. 8
Chuck Maintenance
Note: It is recommended that the push-button chuck be cleaned every 3 - 4 maintenance cycles.
1. Remove bur.
2. Detach handpiece from coupler or delivery system.
3. Fit chuck adapter onto universal cap nozzle of approved lubricant.
4. Insert spray adapter into chuck and dispense spray for 1 - 2 seconds (Fig. 9).
5. Work chuck mechanism to distribute cleaner/lubricant by repeatedly pressing button.
Fig. 9
6. With push-button depressed, insert and remove bur several times to break-up remaining debris.
7. Wipe off excess cleaner/lubricant with a dry, lint-free cloth.