ProMed IRT-100 Instrucciones De Empleo página 22

Termómetro de infrarrojos
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 54
INTRODUCTION
C.
When you hear a long beep, this means that the measurement is finished and you can
read the temperature value on the display.
Caution!
Please ensure that the subject of the measurement (person or surface) and the thermometer
are kept at a stable room temperature for at least 30 minutes before use.
After a continuous measurement, please wait at least two minutes before turning the
thermometer back on.
Do not carry out a measurement immediately after patient care.
Never use the thermometer in a high-temperature environment.
DE
Do not eat, drink or move during the measurement.
Please clean the scanning area and remove dust, hair or sweat before using the
thermometer.
EN
Please remove any water, dust or dirt from the subject of the measurement before
measuring.
Please do not remove the thermometer until you hear the long beep at the end of the
FR
measurement.
If the measuring head gets dirty or you want to pack the thermometer back into the box,
clean it carefully with a cotton cloth soaked in alcohol.
Please attempt to carry out the measurement in the same area since the results in different
IT
areas will be different.
Since a baby's body temperature regulatory function is not perfect, do not carry out a
measurement immediately after a change of environment with a significant difference in
ES
temperature in order to avoid a measurement error.
In the following cases, the use of a conventional thermometer to carry out a measurement
is recommended:
NL
1) The measurement results are higher or lower than expected.
2) For newborn babies who are younger than 100 days old.
3) For children under 3 years of age who are suffering from a defective immune system
RU
or a very high fever.
Accessing the saved measurement values
The thermometer automatically saves 30 measurement values. In order to read the saved
PL
date, proceed as follows.
A. In order to enter memory mode, press the function key (C) for
TR
approx. 1 second with the thermometer turned off and the
symbol "M" will flash in the display.
B.
Now press the function key (C) again and "1" appears in the
AR
display, followed by the last measurement value and the memory
symbol "M" flashes.