Before Use Checklist

• Check that no parts are missing and that all parts are clean.
• To operate the LSU from an external power source, connect to either external AC or
DC power. To operate the LSU from the internal battery, check that the
battery is installed.
• Run Device Test.
• Check that the Serres Vacuum Connector is properly connected to the LSU and to
the canister.
• Check that the lid of the canister is properly sealed: block patient inlet and turn on the
LSU. Vacuum will build up when all parts are assembled correctly.
• Attach appropriate suction catheter if necessary. (Not supplied by Laerdal Medical).

Use

1
1. Unwind the Suction Tubing. Set
the Operating Knob to the required
vacuum level. The LSU will switch on
and start to operate. The Power On
Indicator lights up during operating.
2
2. When suction is complete, set the
Operating Knob to "0" to turn off.
9