How To Measure - DeFelsko PosiTector 200 Manual De Instrucciones

Medidor de espesor de recubrimientos
Ocultar thumbs Ver también para PosiTector 200:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

1. Remove the protective rubber cap from probe.
2. Power-up Gage by pressing the center navigation
3. Apply couplant (pg. 3) to the surface
of the part.
4. Place the probe FLAT on the surface
and press down.
5. Lift the probe when you hear a double
BEEP, - OR - leave probe on the surface
in the same location for continuous
measurements.
6. When completely finished, wipe the probe clean of couplant
then return the Gage to the protective pouch. There is no
need to disconnect the probe from the PosiTector during
storage.
Verify accuracy at the beginning and the end of each shift, or if the
gage is dropped or suspected of giving erroneous readings.
Calibration, Verification and Adjustment
Three steps ensure best accuracy...
1.Calibration - typically done by the manufacturer or a
qualified lab. All probes include a Certificate of Calibration.
2.Verification of Accuracy - as done by the user on known
reference standards such as the included plastic shims, test
block or optional coating thickness standards.
3.Adjustment - to a known thickness, if necessary.
Calibration
Calibration is the controlled and documented process of
measuring traceable calibration standards and verifying that the
results are within the stated accuracy of the Gage. Calibrations
are typically performed by the Gage manufacturer or by a
certified calibration laboratory in a controlled environment using
a documented process.

How to Measure

4
button.