DURING/AFTER POWER FAILURE

 Display of chamber temperature during power failure
By pressing the buzzer stop key (BUZZER) during a "power-failure alarm", the buzzer stops and the
chamber temperature is displayed on the temperature display for 5 seconds.
Check the chamber temperature as appropriate by pressing the buzzer stop key (BUZZER) during
power failure.
<Important>
The battery for the power-failure alarm has a limited lifespan. Replace the battery every 3 years. The
alarm may not be activated if the power fails and the battery has not been replaced regularly. Contact
our sales representative or agent to arrange replacement of the battery.
 Operation check after recovery from power failure
After recovery from a power failure, the operation will resume automatically with the same settings as
before the power failure.
There is no need to re-set.
However, always check the operational
status after recovery.
The settings are memorized during power failure by nonvolatile memory.
<Important>
Upon recovery from a power failure, check that the unit restarts operation properly. Also check that
there are no changes to the settings.
20