REMOTE ALARM TERMINAL
The alarm is relayed to a remote location when a remote alarm device (commercially available) is
connected to the remote alarm terminal. Installation of a remote alarm device is recommended when the
unit is installed in an unattended location so that the operator of the unit is notified of the alarm.
Contact our sales representative or agent to arrange the installation of a remote alarm device.
Location of remote alarm terminal: lower right side
Allowable contact capacity: DC 30 V・2 A
Contact output:
Connection
Between COM. and N.O.
Between COM. and N.C.
Use a twisted shielded wire for the connection.
Type: UL 2343, UL 2448, UL 2464, UL 2552, UL 2623
Length: 30 m max.
The remote alarm terminal will be in alarm mode when the power supply cord is unplugged or the power
switch is turned off because this will be interpreted as a power failure.
The operation of the remote alarm terminal is not conjunction with the alarm buzzer. Therefore, the
operation of the remote alarm terminal cannot be canceled by pressing the alarm buzzer stop key
(BUZZER).
Normal status
Abnormal status
open
closed
closed
open
REMOTE ALARM
MAX DC30V 2A
N.C. COM N.O.
27