This medical product has a temperature resistance of up to max. 135°C (275°F).
Only apply the following sterilization process:
- Autoclave with a triple pre-vacuum (recommended): minimum of 5 minutes at a minimum of 134°C (273°F)
Pay close attention to the manufacturer´s instruction.
6.2 Validation of the processing
i
The validation of the cleaning and of the disinfection was carried out with the device Miele® G 7835 in the programme Vario TD with the
cleanser Neodisher® MediClean Forte.
Adjustment:
- Pre-rinsing for 1 minute
- Cleaning at 55°C (131°F) for 5 minutes with a dosage of 0.5% Neodisher® MediClean Forte
- Neutralization
- Washing down
- Disinfection at 90°C (194°F) for 5 minutes
The sterilization was validated with the device Tuttnauer® EHS 3870 in a fractional pre-vacuum process at 134°C (273°F) for 5 minutes.
Please note, that the reprocessing procedure performed in your dental office must be validated also.
6.3 Maintenance & service
CAUTION
Inappropriate maintenance or service may result in dysfunctions and/or deterioration.
→ Always perform proper maintenance and service.
CAUTION
7 Removal of dysfunctions
7.1 Cleaning of the spray nozzle
WARNING
Damage to pulpal or teeth
A too low quantity of spray water may result in an overheating of the instrument and can lead to resulting injuries.
WARNING
→ Check the canals of the spray water regularly and, in case of blocking, clean the spray nozzle with the enclosed pin.
Carefully remove any blockages of the canals of the spray water with the enclosed pin.
7.2 Storage
Store the cleaned and sterilized instruments in a dry, dark and cool place, sheltered from dust and germs.
Please pay attention to the date of expiry of the sterilization liquid.