5.3.2 Mechanical cleaning and disinfection of the interior and the exterior
VELOCE recommends thermal disinfectors that fulfill the EN ISO 15883-1 requirements and are approved by the manufacturer for the
straight- and contra angle, as well as for high speed handpieces. The cleaning should be carried out at a minimum of 55°C (131°F) and
for at least 5 minutes, the disinfection at a minimum of 90°C (194°F) and for at least 5 minutes (for an A0-value > 3,000). For the
cleaning a mild alkaline cleaning agent with a pH of 9 to 11 is recommended, like e.g. Neodisher® MediClean Forte. For the validation of
the process see chapter 6.2 Validation of the processing.
F For the internal cleaning the adaptors specified by the manufacturer must be used.
F Repeat the process, if visible contaminations still exist after the treatment in a thermal disinfector.
F Before use the instrument must be dry and free of residues.
F To prevent any kind of impairments of the instrument, make sure after every cycle that the instrument is dry on the inside and the
outside.
F Directly afterwards use the VELOCE Premium Service Oil to lubricate the instrument.
5.4 Process of drying
5.4.1 Manual drying
 Blow on the instrument with compressed air from the inside and the outside until no visible drops are left.
5.4.2 Mechanical drying
The drying process normally is part of the cleaning process of your thermal disinfector. Please follow the instructions for your thermal disinfector.
 In case of moisture remaining on the instrument after the cleaning process, blow with compressed air until completely dry.
5.5 Maintenance with the Proclinic service oil
5.5.1 Cleaning and maintaining the spindle
Before each sterilization, please follow these steps in the declared order.
F Please insert the enclosed dental brush into the spindle and carefully remove any kind of debris or residue by circular movements with the
brush. Repeat this process several times and clean the brush with water.
F Place the enclosed lubrication tool VLT1014 on the Proclinic service oil can and spray for approximately 2 seconds.
i
If very unclean, repeat this process several times.
5.5.2 Lubrication of the canal for drive air
VELOCE recommends lubricating and maintaining the instrument after each mechanical cleaning and before each sterilization process.
 For the maintenance of the cartridge screw the lubrication tool (included in the consignment) onto the lubrication can.
 Insert the lubrication tool into the lower end of the instrument and spray into the canal of the driving air for approximately 2 seconds.
 Connect the instrument to your treatment unit and let it run for 15 to 20 seconds, so that the surplus lubricant can escape from the instrument.
6 Sterilization
i
The sterilization bag has to be big enough for the instrument so that the instrument fits without stretching the bag. Concerning quality
and application, the bag has to fulfill all the valid norms and has to be suitable for the sterilization process.
 Shrink-wrap each instrument individually into a sterilization bag.
6.1 Sterilization with a steam sterilizer (Autoclave) according to EN 13060/ISO 17665-1
CAUTION
Damages due to inappropriate maintenance or service may result in dysfunctions and/or deterioration.
Reduced lifetime of the product
CAUTION
→ Lubricate the instrument with the VELOCE Premium Service Oil before each sterilization.
CAUTION
Contact corrosion caused by moisture
Damages of the product
CAUTION
→After sterilization, immediately remove the instrument from the autoclave.