9.
Place patient on the Low Air Loss Mattress. (NOTE: The control unit
can inflate the mattress with the patient laying on it.) To assist in-
patient ingress/egress as well as normal nursing procedure, the Low Air
Loss Mattress can be inflated to maximum by fully rotating the comfort
adjust dial in a clockwise direction.
NOTE: To reduce the risk of skin break down, MAX Inflate should not
be maintained for greater than 30 minutes. At MAX, the inflation
pressure will not exceed 32 mmHg.
10. For proper inflation and patient comfort, adjust pressure by rotating dial
initially to MAX.
sequentially reduce the pressure setting by turning knob counter-
clockwise until approximately four-finger clearance exists between the
underlying foam and the patient's sacral area.
11. For mattress deflation when CPR is needed, disconnect the hose
connector from the controller. Mattress will be fully deflated in less
than 15 seconds.
*PLEASE READ ALL SAFETY INSTRUCTIONS &
WARNINGS ON PAGES 5 & 6 BEFORE USE.
After maximum inflation has been attained,
8