1.8. WATER FLOW DIAGRAM

1.9. SHOOTING TROUBLE

The power cord is connected to the mains,
the electrical switch S is ON, and the red
light on the switch does not turn on.
The water still is on, water is coming out of
drain D and the green L1 light is off.
The water still is on and the water tank is
not at its maximum capacity, no water
circulates.
If the water level in the boiler is below the
temperature sensor and the sensor triggers
Note: It will not be allowed to repair any water still that is not properly cleaned.
Problem
Measure and check that the mains voltage is 230 V.
Check the status of the switch.
Follow this order:
Remove the thermostat protector T and press inwards.
Check level switches Ne and Nd.
Check contactor C1.
Check heaters R1 and R2.
Replace the electrovalve E.
Press the button REARME. (see fig.2.)