Guidance and manufacturer's declaration – electromagnetic Emissions
The SUPERVEGA EVO SUCTION UNIT is intended for use in the electromagnetic environment specified below. The
customers or the user of the SUPERVEGA EVO SUCTION UNIT should make sure that it's used in such an environment.
Emissions test
Irradiated / Conducted
emissions CISPR11
Irradiated / Conducted
emissions CISPR11
Harmonic emissions
EN 61000-3-2
Voltage fluctuations / flicker
emissions EN 61000-3-3
Guidance and manufacturer's declaration – Immunity Emissions
The SUPERVEGA EVO SUCTION UNIT is intended for use in the electromagnetic environment specified below. The
customers or the user of the SUPERVEGA EVO SUCTION UNIT should make sure that it's used in such an environment.
Immunity Test
Level indicated by the
EN 60601-1-2
Electrostatic
+/-8kV on contact
discharge (ESD)
+/-15kV in air
EN 61000-4-2
Electrical fast
± 2kV power supply
transient / burst
lines
EN 61000-4-4
± 1kV for input /
output lines
Surge
± 1kV differential mode
EN 61000-4-5
+/-2 kV ordinary mode
Loss of voltage, brief
<5% U
voltage interruptions
for 0,5 cycle
and variations
40% U
EN 61000-4-11
for 5 cycle
70% U
for 25 cycle
<5% U
for 5 sec
Magnetic field with
30A/m
network frequency
(50/60 HZ)
EN 61000-4-8
Note U
is the value of the power supply voltage
T
ENGLISH
Compliance
Electromagnetic environment - guidance
Group 1
The SUPERVEGA EVO SUCTION UNIT only used RF energy only
for its internal functioning. Therefore, its RF emissions are very low
and are not cause interference in proximity of any Electronic appli-
ances.
Class [B]
The SUPERVEGA EVO SUCTION UNIT can be used in all environ-
ments, including domestic and those connected directly to the public
mains distribution that supplies power to environments used for do-
Class [A]
mestic scopes.
Complies
Compliance Level
The device doesn't
change its state
The device doesn't
change its state
The device doesn't
change its state
(>95% dip U
)
- -
T
T
(60% dip U
)
T
T
(30% dip U
)
T
T
(>95% dip U
)
T
T
The device doesn't
change its state
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic ma-
terial, the relative humidity should be at least
30%.
Mains power quality should be that of a typical
commercial environment or hospital.
Mains power quality should be that of a typical
commercial environment or hospital.
Mains power quality should be that of a typi-
cal commercial environment or hospital If the
user of the SUPERVEGA EVO SUCTION
UNIT request that the appliance operates
continuously, the use of a continuity unit is
recommended.
The power frequency magnetic field should
be measured in the intended installation lo-
cation to make sure that it's sufficiently low.