Output Power - DJO Chattanooga Intelect Mobile 2 Guia De Inicio Rapido

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

I N T E L E C T ® M O B I L E 2 Q U I C K S TA RT G U I D E
A P P E N D I X
ULTRASOUND SPECIFICATIONS
Frequency
Duty Cycles
Pulse Repetition Rate
Pulse duration:

OUTPUT POWER

1 cm
US applicator
Frequency
1MHz 3MHz 1MHz 3MHz 1MHz 3MHz
Effective
Radiating Area
1
ERA INTL (cm2)
Max Output
power in
2W
Continuous mode
Max Output
power in Pulsed
3W
mode
Max Amplitude
2W/
in Continuous
cm
2
mode
Max Amplitude in
3W/
Pulsed mode
cm
2
(*) An error of + 0.25 W can be measured with 1cm2
US applicator, pulse mode 100Hz at 10% or 20% Duty
Cycle.
18
1 MHz; 3 MHz
10%, 20%, 50%, Continuous
16, 48, or 100 Hz
1 -31.25 ms
- Max (ON): 31.25 ms
- Min (OFF): 5ms
2cm
5cm
2
2
0.9
1.5
1
2.5
1.8W
3W
2W
5W
2.7W
4.5W
3W
7.5W
(*)
2W/
2W/
2W/
2W/
cm
cm
cm
cm
2
2
2
2
3W/
3W/
3W/
3W/
cm
cm
cm
cm
2
2
2
2
Unless otherwise specified, ultrasound
controls accuracy is:
Peak to Average Ratio:
Beam Nonuniformity Ratio
Beam Type
Treatment Time
GENERAL SYSTEM OPERATING AND
STORAGE TEMPERATURE
2
Operating Conditions
The device will meet its requirement under the
2.7
following conditions:
Temperature:
Relative Humidity:
5.4W
Atmospheric Pressure:
Transport and Storage Conditions
8.1W
The device will remain in proper condition under the
2W/
following conditions:
cm
2
Temperature:
Relative Humidity:
3W/
Atmospheric Pressure:
cm
2
Time required for the Intelect Mobile 2 to warm from
the minimum storage temperature between uses until
the Intelect Mobile 2 is ready for its INTENDED USE
when the ambient temperature is 20 °C: 5h
± 20 %.
1:1, at 50% Duty Cycle
4:1, at 20% Duty Cycle
9:1, at 10% Duty Cycle
<5:1
Collimating
1 to 30 min
5°C to 40°C
15% to 90%
70kPa to 106kPa
-20°C to 60°C
10% to 90%
50kPa to 106kPa
Tabla de contenido
loading

Tabla de contenido