Alarms Description - Cattani TURBO-SMART A Manual De Instrucciones

Aspirador universal
Ocultar thumbs Ver también para TURBO-SMART A:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

• Alarms description

Alarm code
AC80 AC80-07
00
0
Microcontroller memory alarm
I14
2
Amalgam level > 95%
I15
3
Amalgam level > 100%
I00
32
Microcontroller memory alarm
I01
33
Short-circuit due to one of the two motors
I02
34
Short-circuit before the motors' starting
I03
35
Condensers have not been charged
I04
36
Overtemperature (temperature has exceeded
the max. allowed limit)
I05
37
Blower overcurrent (the blower has
exceeded the current limit)
I07
39
Condensers overvoltage (the condensers
have exceeded the max. voltage limit)
S08
40
Centrifuge short-circuit
S09
41
PC-Board short-circuit on the centrifuge
output
S10
42
Centrifuge instant overcurrent (the centrifuge
has exceeded the current limit)
S11
43
Centrifuge time delayed overcurrent (the
centrifuge has exceeded the current limit)
I16
48
The vacuum detecting pipe is disconnected
I17
49
Centrifuge thermal switch switched-on
DESCRIPTION
SOLUTION
Contact the technician
Replace the amalgam container as soon as
possible
Replace the amalgam container
Contact the technician
Identify where the short-circuit is coming from
and eliminate it
Contact the technician (the PC-Board is likely to
be damaged)
Contact the technician (replace the PC-Board)
Ventilate the plant room
Check the operation of the blower (seized or
excessive friction)
Check the mains voltage (max. 260V)
Eliminate the short-circuit
Replace the PC-Board
Eliminate siphons in the piping or check the
centrifuge operation (seized or excessive friction)
Eliminate siphons in the piping or check the
centrifuge operation (seized or excessive friction)
Connect the vacuum detecting pipe to the
centrifuge
Ventilate the room
35

Hide quick links:

Tabla de contenido
loading

Este manual también es adecuado para:

Turbo-smart b

Tabla de contenido