Do a DPD1 test on the water sample taken from the Tap of the probe holder. Push and quickly release CAL button 2
times and then push + button and confirm by CAL.Scroll the menu by + and – buttons till the display shows "Cl
Calibration" Push CAL to enter the submenu and then the button + to do the GAIN calibration of the probe. If it is
necessary you can modify the value shown in the second row of the display by buttons + or - till the value is equal to
that of DPD1 tester. Push CAL to confirm. If the display shows the message "Calibration Impossible" the probe
calibration is not saved. Read the paragraph named "MESSAGES FROM THE PUMP". If no error message appears
the calibration has been performed.
Codice 1075262
GAIN CALIBRATION
It is recommended to do the gain
calibration after at least 2 or 3 hours of
optimal
working
conditions,
alternatively it is possible to do a first
gain calibration after having installed
the pump and the probe in the plant
and then do the definite calibration
after 24 hours of work of the probe.
After having stabilized the flow and the pH
and
having
stabilized
concentration at the optimal and desired
value, wait for the stabilization of the
reading on the display.
pag. 30 a 56
or
the
chlorine
Ver.01 – 19/06/20