Hazardous area
Zone 0, 1, 2, 20, 21, 22
Terminal (31,32)
Supply:
Voltage
19.2 – 31.2 VDC
Power max.
3.5 W
Terminal (33,34)
Status Relay:
Non Hazardous location
Voltage max.
125 VAC / 110 VDC
Power max.
62.5 VA / 32 W
Current max.
0.5 AAC / 0.3 ADC
Revision date:
2011-11-20
9203 - Product Version 9203-002
Non Hazardous area
or Zone 2
-20 ≤ Ta ≤ 60ºC
4501
44
43
42
41
54
53
52
51
9203
91 92 93 94 95
Power
Rail
Terminal (11,12 and 13,14)
Input:
Voltage
Trig: NPN
Trig: PNP
Zone 2 installation
32 VAC / 32 VDC
16 VA / 32 W
0.5 AAC / 1 ADC
Version Revision
V5 R0
34
33
32
31
14
13
12
11
Supply / Input
(terminal 11,12,13,14)
(terminal 31,32,33,34)
(terminal 91,92,93,94,95)
U
: 253V, max. 400Hz
m
max 28VDC
Low < 2 V, High > 4 V
Low < 8 V, High > 10 V
Prepared by:
PB
9203QI01
Page:
2/3
21