PASSWORD = 0 (PASSWORD not programmed)
pH STABILITY= NO (deactivated)
START DELAY: 0 minutes
pH:
OPERATION TYPE: Proportional
SETPOINT: 7.3 pH
PROPORTIONALITY:1pH
REGULATION DIRECTION: Acid (pH - )
FLOW RATE: 100%
DELAY ON: 3 seconds (0-999sec)
ALARM TIME : 0unit (disabled)
DEFAULT SETTINGS RESTORE
MENU TREE
Push and quickly release CAL button during the functioning to enter into the programming menu. Push + or - button to
navigate through the menu. Push the CAL button to enter the submenus.

LANGUAGE SELECTION

DELAY START

WHAT IS IT? The delay start is the time in minutes (0-99min) after the switching on that DUAL waits before dosing the
chemical product or activating the Rx power plug (on DUAL pH/REG). During this period of time the display shows the
message "Del" on the second line alternativenly with the measure (see figure here above). During this time of delay DUAL
can't dose or activate the power plug (on DUAL pH/REG) but it is possible to enter the menu and modify the parameters
and calibrations.
Rev 1.0
DEFAULT SETTINGS
Rx:
OPERATION TYPE: Proportional
SETPOINT: 730mV
PROPORTIONALITY:100mV
REGULATION DIRECTION: Oxidizing
FLOW RATE: 100%
DELAY ON: 3 seconds (0-999sec)
ALARM TIME : 0unit (disabled)
Push and quickly release CAL button to enter programming and then push
+ and – buttons to select "Factory Settings Restore"
Push CAL to confirm the choice. The pump goes back to measure
restoring in memory the default settings and calibrations parameters.
Push and quickly release CAL button and then push + and – buttons to select
the language choice "Language Selection". Push CAL to enter the language
submenu and select " English" by + or - .
Push CAL to confirm the choice and ESC to go back in measuring.
DELAY START EDITING
Push and quickly release CAL button and scroll the menu using + and – buttons
till the display shows "Delay Start". Push CAL to enter the submenu and then,
by pushing + and – button, choose the value in minutes (0-99). Push CAL to
confirm and ESC to go back in measuring.
Please note that the change will take effect from the next power on of
DUAL!