• Gently remove fuse and install new fuse into the holder.
• Always use a UL recognized fuse of the proper size and value: 800mA/1000V (6.3 x
32mm) fast blow for the μA/mA ranges and 10A/1000V(10 x 38mm) fast blow for the
10A range.
• Install the back cover and tighten the screw.
WARNING: To avoid electric shock, do not operate meter until the battery/fuse cover
is securely fastened to the meter.
6. Specifications
Accuracy is stated at 65°F to 83°F(18°C to 28°C), less than 70% 1- relative humidity
Function
Range
Resolution
6V
1mV
60V
10mV
AC Voltage
600V
0.1V
1000V
1V
Input Protection: 1000V AC RMS or 1000V DC
Accurary specified from 5% to 100% of the measuring rang.
AC voltage accuracy bandwidth: 45Hz to 1000Hz
Distorted, pulsed, triangle or trapezia waveform.
Accuracy: ±(10%rdg + 10dgt)
6V
1mV
Low Z
60V
10mV
AC Voltage
600V
0.1V
Input Protection: 600V AC RMS or 600V DC
Accurary specified from 5% to 100% of the measuring rang.
Input Impedance: Approx. 3kΩ
600mV
0.1mV
6V
1mV
DC Voltage
60V
10mV
600V
0.1V
1000V
1V
Input Protection: 1000V AC RMS or 1000V DC
12
Accuracy ±(% of reading + digits)
±(0.8% + 8)
±(1.0% + 3)
±(3.0% + 40)
±(0.5% + 8)
±(0.5% + 5)
±(0.8% + 3)
13/12/2017 Version No. 001
Function
Range
Resolution
600mV
0.1mV
6V
1mV
Low Z
DC Voltage
60V
10mV
600V
0.1V
Input Protection: 600V AC RMS or 600V DC
Input Impedance: Approx. 3kΩ
9.999Hz
0.001Hz
99.99Hz
0.01Hz
Frequency
(electrical)
999.9Hz
0.1Hz
9.999kHz
1Hz
Input Protection: 600V AC RMS or 600V DC
Sensitivity: <8V RMS
Duty Cycle
20.0% to 80.0% 0 .1%
Input Protection: 600V AC RMS or 600V DC
Pulse Width: 0.1 to 100mS
Frequency Range: 5Hz to 10kHz
Sensitivity: >8V RMS
600μA
0.1μA
6000μA
1μA
AC Current
60mA
10μA
600mA
0.1mA
10A
10mA
Overload Protection: μA, mA ranges: 800mA/1000V Fuse
10A range: 10A/1000V Fuse
Accurary specified from 5% to 100% of the measuring rang.
AC current bandwidth: 45 to 500Hz
Distorted, pulsed, triangle or trapezia waveform.
Accuracy: ±(10%rdg + 10dgt)
13/12/2017 Version No. 001
True RMS Digital Multimeter/ English
Accuracy ±(% of reading + digits)
±(3.0% + 40)
±(1.0% + 5)
±(1.2% + 2)
±(1.5% + 3)
±(2.0% + 5)