• If the color remains purple, or a color similar to purple, the patient may
not have a patent airway or the endotracheal tube may not be positioned
correctly. This requires immediate attention. Employ your institution's
protocol to verify proper tube placement and adequate gas exchange.
• Continue to monitor the color of the CO
entire time the device is in use.
DO NOT exceed 24 hours of use.
• During the course of ventilation, if the FLOCAP returns to and
remains purple or beige colored, this indicates insufficient exhaled
CO
. This requires immediate attention.
2
4. Not intended to be a quantitative indicator. Follow your institution's
guidelines for breath frequency and tidal volume during resuscitation
and for avoiding auto-PEEP during resuscitation. At the end of
exhalation, the flow indicator's momentum will cause it to spin briefly
after exhalation has ceased.
3.0 COLOR & FLOW INDICATOR INTERPRETATION
Purple (0%) .................................................................No carbon dioxide detected
Beige (1-2%) ...................................................Exposed to CO
Yellow (5%) ............................................................ Exposed to 5% or greater CO
Spinning * .........................................................................................Patient is exhaling
Not Spinning ......................................................................Exhalation flow < 3 LPM
* At the end of exhalation, the flow indicator's momentum will cause it to
spin briefly after exhalation has ceased.

4.0 MECHANICAL SPECIFICATIONS

Internal Volume: ..................................................................................................< 25 ml
Pressure Drop according to ISO 9360-1:
at 30 LPM ...................................................................................................... 0.7 cmH
at 60 LPM .......................................................................................................2.7 cmH
at 90 LPM ...................................................................................................... 5.7 cmH
Leak rate according to ISO 9360-1: ..................................................... 0.0 ml/min
Compliance according to ISO 9360-1: .............................................0.44 ml/kPa
Device Weight: ................................................................................................23 grams
Connector ports according to ISO 5356-1:
Patient end ...............................................................................15mm I.D./22mm O.D.
Ventilator end ............................................................................................... 15mm O.D.
6
866.4.Maxtec www.maxtec.com
indicator throughout the
2
, approx. 1% to 2%
2
2
O
2
O
2
O
2