troubleshooting
Problem
Probable Cause
The display
An error has occurred in the
shows a device
device and requires servicing.
fault.
Symptoms of
1. Air filter may be occluded.
OSA have
recurred.
2. Sleep apnea condition has
Your skin
1. Headgear is too tight or
becomes
irritated where
the mask
2. Your mask may not be
touches your
face.
Dryness of
Inadequate humidity.
throat or nose.
Condensation of
1. Your humidification setting is
water is
collecting in the
2. The room temperature
hose causing a
gurgling noise.
Air from flow
1. The air filter(s) is/are dirty.
generator seems
to be too warm.
2. The air inlet port is blocked.
3. The room temperature is too
4. The IntelliPAP is located near
Nasal, sinus or
You may have a reaction to air
ear pain,
flow pressure.
possibly with
runny nose.
changed.
improperly adjusted.
properly sized or is not the
most appropriate shape for
you.
too high.
fluctuates from higher to
lower levels through the
night.
high.
a heater.
Remedy
Contact your home care provider
for service.
1. Clean or replace the air filter
and relocate device away from
drapes and other dusty surfaces.
2. Contact your physician.
1. Loosen the headgear to reduce
the contact pressure on your
face.
2. Contact your physician or home
care provider.
1. Add a DeVilbiss DV5 Heated
Humidifier.
2. Increase the heater setting on
the humidifier.
Reduce the humidifier heater
setting and/or increase the
temperature of the room.
1. Refer to the Cleaning section for
instructions on filter
maintenance.
2. Unblock the air inlet.
3. Reduce the room temperature.
4. Locate the device away from
the heat source so the air
being drawn into the device is
at room temperature.
If the problem persists contact
your home care provider.
Discontinue use and contact your
physician.
A-DV51
EN
21