MANUFACTURER'S GUIDE AND DECLARATION - ELECTROMAGNETIC IMMUNITY
The clinical thermometer JC-132G can be used in the following electromagnetic environment. The customer or
user of the clinical thermometer JC-132G must make sure that the device is used in said environment.
EMISSION TEST
Electrostatic discharges (ESD)
IEC/EN61000-4-2
Magnetic field
IEC/EN 61000-4-8
Irradiated immunity
IEC/EN61000-4-6
All the indications, specifications and drawings are based on the latest
information available when this manual was printed. They can be subject to
variations or changes with no notice required.
COMPLIANCE
± 6kV in contact
± 8kV in the air
3 A/m
3 V/m 80 MHz at 2.5 GHz (in
appliances that are not life
equipment)
12
GUIDE TO THE ELECTROMAGNETIC
ENVIRONMENT
The floors should be made of
wood, cement or baked clay. If
the floors are covered in synthetic
material, relative humidity should
not exceed 30%.
The magnetic field should
be the typical magnetic field
of a commercial or hospital
environment.