MANUFACTURER GUIDANCE AND DECLARATION – ELECTROMAGNETIC IMMUNITY
Nebulizer JC-114 can be used in the following electromagnetic environment. The customer
or the user of NEBULIZER JC-114 must make sure to use it in this environment.
EMISSION TESTS
Electrostatic
discharge
(ESD)
IEC/EN61000-4-2
Fast
± 2kV power supply
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN61000-4-5
Voltage dips, short
< 5% UT for 0.5 cycles
interruption and
40% UT for 05 cycles
power fluctuations
70% UT for 25 cycles
IEC/EN61000-4-11
Magnetic field
IEC/EN61000-4-8
Conducted
immunities
IEC/EN61000-4-6
(for devices that are
not life-supporting)
Radiated
immunities
IEC/EN61000-4-3
(for devices that are
not life-equipment )
Note: UT is the a.c. mains voltage prior to application of the test level
Specifications and designs are based on the latest information available at the time of printing
and subject to change without notice.
ULTRASONIC NEBULIZER
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT GUIDE
± 6kV in contact
Floor should be concrete, wood or ceramic
± 8kV in air
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment
± 1kV differential
Mains power quality should be that of a
mode
typical commercial or hospital environment
Power supply should be the same as the one
supplied in the commercial and hospital
environment. Whether the user needs a
< 5% UT for 5
continuous use, it will be necessary to
seconds
connect the unit to an UPS (uninterruptible
power supply)
3 A/m
Power frequency magnetic fields should be at
level characteristics of a typical location in a
typical commercial or hospital environment.
3Vrms 150kHz
to 80MHz
3 V/m 80 MHz a
2.5 GHz