GB
JC-114
INSTRUCTIONS FOR A CORRECT USE OF THE PRODUCT
The Joycare Ultrasonic Nebuliser JC-114 can be use in two different way: that is to
say with or without the medicine cup.
Because the device is not sterile, before using go on with cleaning and disinfection
operations indicated in the specific paragraph. Dosage indication and use
limitations provided by disinfectant producer.
Nebulisation without the medicine cup
1. Take the upper cover away
2. Pour 10cc of drug at most, directly into the reservoir
3. Never place the reservoir under the faucet for water filling
4. Once the reservoir has been filled, put the upper cover back on the main body
carefully and insert the accessory you want to use (tube, mask, mouthpiece and
nasal fork).
5. Plug in the adapter and then connect it with the main power source.
6. Switch the appliance on and start the treatment. Be sure that the LED is on. The
blue light indicates a high intensity nebulisation, that is 13 ± 2 minutes at most;
the yellow light, indicates a lower intensity nebulisation, that is 26 ± 2 minutes at
most. Once the drug has ended, the appliance turns itself off automatically.
Before going on with therapy, to verify the correct assembly of each component,
turn on the appliance and control that the nebulized drug goes out under the
shape of vapour, only from the openings of the inserted accessory. Thereby you
can go on with therapy.
7. While you are using the appliance never hold or slope it. Never remove the main
unit.
8. When the nebulisation stops, this would mean the treatment has ended. Now
you can unplug the appliance and start cleaning the unit.
Nebulisation with medicine cup
1. Take the cover away
2. Pour 5~6cc of water into the reservoir
3. Never place the reservoir under the faucet for water filling
4. Put the medication cup into the reservoir. Fill the cup with 10cc drug max. After
having checked the medication cup is in the correct position, put the upper
cover back on the main unit and insert the accessory you want to use (tube,
mask, mouthpiece and nasal fork).
5. Plug in the adapter and then connect to the power source.
6. Switch the nebuliser on and start the treatment. Be sure that the LED is on. The
blue light indicates a high intensity nebulisation, that is 13 ± 2 minutes at most;
the yellow light, indicates a lower intensity nebulisation, that is 26 ± 2 minutes at
ULTRASONIC NEBULIZER