out of the compartment in time. Or the volatilization of the solution
would make the mirror go moldy. Users must pay more attention to the
corrosive sample and liquid easy to volatilize. Any solution remains
in the compartment should be wiped off immediately.
Surface Clean
The cover of the instrument is with paint. Please use wet towel to
wipe off the drips on the surface immediately. Organic solution is
forbidden to be used to clean the cover. Please wipe off the dirt on
the cover timely.
Clean the Cuvettes
After every test or after a solution change, the cuvettes should be
cleaned carefully, or the remains on the surface would cause measuring
error.
4. Spare Parts Replacement
Replace Fuse
Danger! Be sure to switch off the power and unplug the
socket before replacement!
Step 1.
Tools preparation
Prepare a 3×75 Flat Blade screwdriver.
Step 2.
Switch Off the power supply
Switch off the power supply, and unplug the socket.
Step 3.
Take out the Fuse Seat
Take out the fuse seat by the screwdriver.
Step 4.
Replace a new fuse
Pick out the spare fuse (3.15A/250V) and replace it to the working
position.