WL: 500.0nm
Main Menu
Basic Mode
①
Quantitative Mode
②
Wavelength Scan
③
Kinetics Mode
④
DNA/Protein Mode
⑤
Multi-Wavelength Mode
⑥
System Utility
⑦
Cancel
Step 2.
Set Photometric Mode
Press
to set photometric mode. Press
"T%" or "Conc./Factor" and press
"Abs." or "T%", please go to step 5 directly.
Step 3.
Set Concentration Unit
Press
to set concentration unit. Press
followed with
to input the unit.
Step 4.
Set "Factor" or "Standard"
Two methods are under your choice:
Method 1:Input Factor F
Press
to set F. Input the value of F by numeric keypad, press
EN TER
to confirm. Then the F value would display on the screen.
Method 2:Use Standard Sample
Put the reference sample in the light path and calibrate 100%T/0Abs;
Put the standard sample in the light path,press
mark. Input the concentration value of the standard and press
to confirm, then it displays on the screen.
Step 5.
Set Wavelength
GOTO ʎ
Press
to set wavelength, input the value by the numeric keypad
followed with
10:00:00
D2
W
Move Dn
Select
EN TER
pressed to confirm. You can also choose "other"
EN TER
pressed to confirm.
WL: 500.0nm
100.0 %T
Unit
Mode
Factor
,
to choose "Abs.",
EN TER
to confirm. If users choose
,
10:00:00
D2
W
Standard
to choose unit
to start the
EN TER