KONTROLab XVet 28 Manual De Usuario página 33

Rayos x portatil veterinario
Tabla de contenido
Manual ver1.7
Appendix-4 Tabla de errores
ERROR
No.
Código
1
002
THERMAL ERR(INTERLOCK)
2
010
3
011
4
012
5
014
6
018
7
016
8
015
run EP FB or AC POWER LOW ERR
9
017
10
007
11
008
Descripción
standby FIL FB ERROR
standby EP FB ERROR
standby IP FB ERROR
run FIL FB ERR
run IP FB LOW ERR
run IP FB HIGH ERR
run EP FB HIGH ERR
READY SW ERR
EXPOSURE SW ERR
Indicación
Excess from 60℃
Excess from 200mA (10% of max. current, 2A)
Excess from 12KV (10% of max. voltage, 120KV)
Excess from 3.5mA (10% of MAX current, 35mA)
Out of range from reference±10%
Out of range from reference-10%
Out of range from reference+10%
Out of range from reference-10% or AC power line poor
Out of range from reference+10%
READY switch malfunction
EXPOSURE switch malfunction