BeamControl-Master BCM
Calibration procedure
6. Once you have completed steps 1 to 4 and identified dots A2 (reference) and A3
(deviation), use the corresponding screws to level the laser so that dot A3 covers dot
A2. During this process, the air bubble in the level will move out of its central positi-
on and stray to the side.
7. Remove the protective caps from the ends of the levels to
expose the calibration screws. Turn the screws to calibrate the
level (using the SW 2.5 socket wrench included in the scope of
supply). Calibration is complete when the air bubble returns to
the centre of the level.
8. Repeat for the remaining axes.
Technical data
Accuracy
Laser
Output power
Laser wavelength
Beam splitter
Reference beam
Horizontal plano
Speed of rotation
Battery life
Continious operation
Working- / Storage temperature 0° C to 50° C / -10° C to 70° C
Weight
SensoCommander 120 / 310 (optional)
Batteries/Power supply:
SensoCommander 120 /
SensoCommander 310
Battery life
Laser reception range
SensoCommander 120
SensoCommander 310
Working temperature
Storage temperature
18
(Subject to technical alterations)
± 2 mm / 10 m
class 2
< 1 mW
635 nm
20"
vertical
inclination can be adjusted manually
0 ... 120 rpm variable,
550 rpm for receiver
40 h (4 x 1,5 V of type AA Mignon alkaline)
connection socket for external supply
0,8 kg
2 x type AAA /
1 x 9 V block
approx. 70 h
max. 120 m
max. 300 m
-10°C ... + 50°C
-10°C ... + 70°C
ENGLISH
Calibration screws