Introduction
CAUTION!
Prerequisites for operation
Possible consequence: Slight or minor
injuries, material damage
–
The sensor may only be used with
ProMinent type DLG III or DGM flow
gauges. Only in this way can the flow
parameters be maintained
–
There must be a free outlet or at most
a counter pressure of 1 bar at the
outlet of the flow gauge
–
The sensor's voltage supply may not
be interrupted
–
Following longer interruptions to the
voltage supply (> 2 h), allow the
sensor to run-in again and recalibrate
it
1.4
Correct and Proper Use
NOTICE!
Correct and Proper Use
–
The sensor may only be used to
determine and regulate concentra‐
tions of free chlorine
–
The sensor may not be used in con‐
junction with organic chlorine com‐
pounds, such as trichloroisocyanuric
acid, or stabilisers, such as cyanuric
acid
–
Any other uses or modifications to the
system are prohibited
–
The sensor is not a safety component