Power
frequency
(50/60 HZ)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration-electromagnetic immunity
The FORA TN'G BP Blood Pressure Monitoring System is intended for use in the
electromagnetic environment specified below.
The customer or the user of the FORA TN'G BP Blood Pressure Monitoring System
should assure that is used in such and environment.
Immunity
IEC 60601
test
test level
Conducted
3 Vrms
RF
150 KHz
IEC
to 80 MHz
61000-4-6
Radiated
3 V/m
RF
80 MHz to
IEC
2,5 GHz
61000-4-3
36
3 A/m
Compliance
level
Not
applicable
3 V/m
The FORA TN'G BP Blood
Pressure Monitoring System
power frequency magnetic fields
3 A/m
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
Electromagnetic
environment-guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the FORA TN'G BP Blood Pressure
Monitoring System including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).