Guidance and manufacturer's declaration-electromagnetic immunity
The FORA TN'G BP Blood Pressure Monitoring System is intended for use in the
electromagnetic environment specified below.
The customer or the user of the FORA TN'G BP Blood Pressure Monitoring System
should assure that it is used in such an environment.
IEC 60601 test
Immunity test
level
Electrostatic
discharge
+ 6 kV contact
(ESD)
+ 8 kV air
IEC 61000-4-2
Electrical fast
+ 2kV for power
transient/
supply lines
burst
+ 1kV for input/
IEC 61000-4-4
output Lines
+ 1kV line(s) to
Surge IEC
line(s)
61000-4-5
+ 2kV line(s) to
earth
<5% UT (>95%
dip in UT) for
Voltage Dips,
0,5 cycle
short
40% UT(60%
interruptions
dip in UT) for 5
and voltage
cycles
variations on
70% UT(30%
power supply
dip in UT) for 25
input lines IEC
cycles
61000-4-11
<5% UT(>95%
dip in UT) for 5 s
Compliance
Electromagnetic
level
environment-guidance
Floors should be wood, concrete
or ceramic tile. If floors are
+ 6 kV contact
covered with synthetic material,
+ 8 kV air
the relative humidity should be
at least 30%.
Not
Mains power quality should be
applicable
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the FORA TN'G BP
Blood Pressure Monitoring
Not
System requires continued
applicable
operation during power mains
interruptions, it is recommended
that the FORA TN'G BP Blood
Pressure Monitoring System be
powered from an uninterruptible
power supply or a battery.